Newton Science micropipettors are budget-level, air displacement, microlitre pipettors designed for use in a wide range of applications.
General specifications:
- Digital variable volume or fixed volume models available
- Fully autoclavable (at 121˚C and 15psi for 10-15 minutes)
- Variable volume models have one-handed adjustment and 4-digit display with high-contrast figures for easy viewing
- Streamlined tip ejector for smooth tip ejection
- Large handgrip for comfortable operation
Performance data, fixed volume models:
Volume, µl | Accuracy (A, ±%) | Coefficient of variation* (CV, ±%) |
1 | 2.0 | 1.0 |
2 | 2.0 | 1.0 |
5 | 2.0 | 1.0 |
10 | 1.0 | 0.5 |
20 | 0.8 | 0.4 |
25 | 0.8 | 0.4 |
50 | 0.8 | 0.4 |
100 | 0.6 | 0.2 |
200 | 0.6 | 0.2 |
250 | 0.6 | 0.2 |
500 | 0.6 | 0.2 |
1000 | 0.6 | 0.2 |
2000 | 0.6 | 0.2 |
5000 | 0.6 | 0.2 |
10000 | 0.6 | 0.2 |
* Figures quoted are based on using distilled water at 20°C.
Supplied complete with a performance certificate and shelf, or rack, hanger.