For use in applications examining pharmaceutical and biochemical samples, clinical and biological fluids, industrial processes and environmental testing.
- Single channel emission flame photometer for routine determination of Na, K and Ca (Li and Ba also, using accessory filters).
- Digital readout with moveable decimal point
- Automatic air regulation
- Auto-ignition of flame
- Gas cut-off safety feature
- Integral ‘fail-safe’ flame monitoring
- Recorder output
Model | 410 Classic | |
Sensitivity | Minimum concentrations producing a display reading of 100 scale units: Na: 0.5ppm. K: 0.5ppm, Li: 2ppm, Ba: 1000ppm (optional filter), Ca: 5ppm | |
Detection limts | Na, K <0.02ppm at 2 x standard deviation | |
Specificity | <0.5% to reading from interferent of Na, K, Li equal to the concentration of element under test | |
Drift | % | 2 |
Reproducibility | % | 1 |
Display channels | 1 (LED) | |
Result display | Manual estimation | |
Result storage | No | |
Outputs | Fixed analogue | |
Dimensions | mm | 390 x 345 x 510 |
Weight | kg | 9.5 |
Services required | ||
Electrical | 198-264V 50/60Hz | |
Air | Moisture and oil free 6 litres/minute at 1kg/cm2 | |
Gas | High grade propane, butane or propane/butane mixture. Fuel should be free from heavy hydrocarbon deposits and regulated to 30 p.s.i. at the cylinder. Natural gas at 3" to 10" W.G. is also suitable (specifiy correct regulator when ordering) |
With Na, K and Ca filters, fuel and air connection, 1000ppm standard solutions for Na, K and Ca and instructions. Requires but does not include an air compressor of the correct type and gas regulator. For 198-264V 50/60Hz single phase supplies.
Clinical version also available - details on request.